Accuracy Testing of the MAX30100 Sensor Using the Contec MS100 SpO2 Simulator
DOI:
https://doi.org/10.30595/pshms.v9i1.2229Keywords:
MAX30100 sensor, SpO2, Pulse rate, Accuracy, Contec MS100 simulator, Wearable devicesAbstract
The accuracy of pulse oximeters is critical for reliable health monitoring. The MAX30100 sensor is widely used in wearable devices to measure SpO2 and pulse rate. However, its accuracy needs to be assessed against reliable standard devices. The Contec MS100 simulator is used as a benchmark to evaluate the performance of such sensors. A descriptive quantitative study was conducted with a sample of 347 participants. Data were collected from both the MAX30100 sensor and the Contec MS100 simulator to measure SpO2 and pulse rate. The Bland-Altman method and Kruskal-Wallis test were used for statistical analysis to assess the accuracy and agreement between the two devices. The study found that the MAX30100 sensor provided reliable measurements for both SpO2 and pulse rate. The mean difference between the MAX30100 and Contec MS100 was 0.45% for SpO2 and 0.5 bpm for pulse rate. The Bland-Altman analysis showed acceptable limits of agreement for both parameters. However, minor discrepancies were observed. The study found that the MAX30100 sensor provided reliable measurements for both SpO2 and pulse rate. The mean difference between the MAX30100 and Contec MS100 was 0.45% for SpO2 and 0.5 bpm for pulse rate. The Bland-Altman analysis showed acceptable limits of agreement for both parameters. However, minor discrepancies were observed.
References
[1] Ramadhani S, et al. Accuracy of non-invasive sensors in clinical settings. J Health Sci. 2023;21(3):134-
139.
[2] Siregar Z, et al. Performance evaluation of the MAX30100 sensor for pulse oximetry. Med Devices.
2022;13(2):110-115.
[3] Lovibond PF, Lovibond SH. Manual for the Depression Anxiety Stress Scales. 2nd ed. Sydney: Psychology
Foundation; 1995.
[4] Buysse DJ, Reynolds CF, Monk TH, Berman SR, Kupfer DJ. The Pittsburgh Sleep Quality Index: A new
instrument for psychiatric practice and research. Psychiatry Research. 1989;28(2):193-213.
[5] Johnson L, et al. Evaluation of the MAX30100 sensor: Accuracy and limitations in clinical settings. J Med
Technol. 2022;18(1):75-82.
[6] Azzura S, et al. The impact of sleep quality on hormonal balance and menstrual cycle regulation. J Sleep
Res. 2023;12(4):210-220.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Proceedings Series on Health & Medical Sciences

This work is licensed under a Creative Commons Attribution 4.0 International License.